The Data of Lyophilization

Aug 24, 2022 Leave a message


  •  Pre-freezing speed

      The pre-freezing speed cannot be controlled by most machines, so the pre-freezing speed can only be determined by the pre-freezing temperature and the packing time. If the pre-freezing speed is required to be fast, the freeze-drying box will be lowered to the lower temperature first, and then the product will be released. if the rate of pre-freezing is required to be slow, the freeze-drying box should be cooled down after the product is in the box.

  • Minimum temperature for pre-freezing

      This temperature depends on the eutectic temperature of the product, and the minimum pre-freezing temperature should be lower than the eutectic temperature of the product.

  • Pre-freezing time

      The product is loaded with a large amount, and the bottom of the container used is thick and uneven. The product is not directly placed on the freeze-drying box for freeze-drying. The freeze-drying box has poor freezer capacity. Machines with large temperature difference between parts require longer pre-freezing time. In order to freeze every bottle of product in the box, it is generally required to keep the temperature of the sample for 1-2 hours after the temperature reaches the predetermined minimum temperature.

  • Condenser cooling time

      The condenser is required to cool down at the end of the pre-freezing period, before the pre-freezing has ended, and before vacuuming. How much time before depends on the cooling performance of the condenser machine. It is required that the temperature of the condenser should reach about -40 ℃ when the vacuum is drawn after the pre-freezing. A good machine generally starts to cool down half an hour before. 

     The cooling of the condenser is usually continued from the beginning until the freeze-drying is completed. The temperature should always be below -40°C.

  •  Vacuum time

      The end of pre-freezing is the time to start vacuuming, and it is required that the vacuum degree can reach 1×10-4 mbar in about half an hour.

      At the same time as vacuuming, the vacuum valve between the condensers of the freeze-drying box is opened, and the vacuum pump and the vacuum valve are also opened until the freeze-drying is completed.

  • Pre-freezing end time

      The end of the pre-freezing is the operation of the freeze-drying box freezer. Usually, the operation of the freezer is stopped at the same time as the vacuum is pumped or when the vacuum is pumped to the specified requirements.

  • Start heating time

      It is generally considered the time to start heating (in fact, the sublimation starts when the vacuum is drawn). The heating is started after the vacuum degree reaches 1 × 10-1 mbar (close to 1 × 10-1 mm Hg), and some freeze dryers use the vacuum relay to automatically turn on the heating, that is, the air degree reaches 1 × 10-1 mbar When the temperature is reached, heating will start automatically; some freeze dryers start heating half an hour after vacuuming, and the vacuum degree has reached 1 × 10-1 mbar or even higher.

  • Vacuum alarm working time

      Since the degree of vacuum is extremely important for sublimation, the new freeze dryers are equipped with vacuum alarm devices. The working time of the vacuum alarm device is from the start of heating to the use of calibrated leaks, or from the beginning to the end of freeze-drying.

      Once the vacuum degree drops during the sublimation process and a vacuum alarm occurs, on the one hand, an alarm signal is issued, and on the other hand, the heating of the freeze-drying box is automatically cut off. At the same time, the freezer of the freeze-drying box is also activated to cool the product to protect the product from melting.

  • Working time to correct leaks

      The purpose of correcting the leak is to improve the heat transfer in the freeze-drying box and is usually used during the second stage of work, continuing to restore the high vacuum state. The length of use time is determined by the variety of products, the amount of filling and the value of the vacuum degree set.

  • Maximum allowable temperature for product heating

      The maximum permissible temperature of the board layer heating is determined according to the product. During the sublimation, the heating temperature of the board layer can exceed the maximum permissible temperature of the product because the product is still in the low temperature stage. The layer temperature needs to drop to match the maximum allowable temperature of the product. Due to the temperature difference of heat transfer, the temperature of the sheet layer can be slightly higher than the maximum allowable temperature of the product.

  • Total time for lyophilization

      The total time for lyophilization is the pre-freeze time, plus the sublimation time and the time for the second stage work. The total time is determined, and the lyophilization end time is also determined.

      This time depends on the variety of the product, the variety of the bottle, the packing method, the filling amount, the machine performance, etc. Generally, the freezing time is longer, about 18-24 hours, and some special products need a few days.